xue ye zhong san lv yi xi de dai bo ji zuo ding kong qi xiang se pu ce ding fa
Author(s): QIN Wenhua, LIU Jia, HAN Huayun
Pages: 499-
500
Year: 2009
Issue:
8
Journal: Chinese Journal of Industrial Hygiene and Occupational Diseases
Keyword: 色谱法; 气相; 血液; 三氯乙烯;
Abstract: 目的 简化血液中三氯乙烯测定方法中对血液样品繁杂的前处理步骤,建立一种简单的测定方法.方法 用带捕集阱顶空仪进样,气相色谱法经Elite-XLB柱分离后用电子捕获检测器检测.结果 三氯乙烯浓度为0.5~80.0 μg/L呈良好的线性关系,相关系数r=0.999 2,同归方程为Y=4150.72X-10 127.59.检出限为0.5 μg/L.在线性范围内,对高、中、低浓度测定的相对标准偏差分别为4.05%、7.28%和7.44%.在实际样品中进行加标回收试验,回收率为96.83%~105.00%.干扰试验证明,其他卤代烃元素不干扰三氯乙烯测定.结论 该方法操作简便,灵敏度较高,完全能满足血液中三氯乙烯检测的需要.