A clinical accuracy evaluation of an electronic apex locator
Author(s): ZHANG Yu-xing, LIN Nan-yan, ZHANG Wei-hong
Pages: 39-
40
Year: 2005
Issue:
7
Journal: CHINESE JOURNAL OF MEDICINE
Keyword: 根管长度; 电测法; 准确性;
Abstract: 目的探讨根尖炎症及根管干、湿状态对电测法测量根管工作长度准确性的影响.方法临床上随机选择拟做根管治疗的120颗牙共140个根管进行研究,其中无根尖炎和有根尖炎者分别为56、84个根管.常规根管清理后,采用ROOT ZX根管长度电测仪在干、湿两种根管状态下进行根尖位置的探测,根管充填后,根据X线片影像确定根管长度电测法的准确性. 结果根管充填后X线片显示恰填者104个(74.29%)根管,欠填者6个(4.28%)根管,超填者30个(21.43%)根管.干燥的根管较湿润的根管更易测得准确的根管工作长度(P<0.01);有、无根尖炎其测量结果准确性无显著性差异(P>0.05).结论根管长度电测法是一种准确度高、操作简便且实用的确定根管工作长度的方法.根尖炎症对其测量根管工作长度的准确性无明显影响,干燥的根管较湿润的根管更易测得准确的根管工作长度.