chi xu qi dao nei zheng ya tong qi zhi liao zu sai xing shui mian hu xi zan ting zong he zheng de ya li diao ding
Author(s): 
Pages: 607-608
Year: Issue:  10
Journal: MODERN PRACTICAL MEDICINE

Keyword:  睡眠呼吸暂停低通气综合征持续正压通气压力;
Abstract: 目的观察自动压力调节系统与传统的手动压力调节对阻塞性睡眠呼吸暂停综合征(OSAHS)患者持续气道内正压通气(CPAP)治疗压力调定的效果. 方法 44例OSAHS患者,随机分为2组,对照组24例用手动法、治疗组20例用自动压力调节系统分别作压力调定,并以此为标准作CPAP治疗,监测相关的呼吸指标. 结果经手动调定的压力值为5.0~13.5cmH2O,平均10.10±2.38cmH2O;自调动定的压力值为5.0~12.4 cmH2O,平均7.89±2.08cmH2O.两组治疗后的呼吸暂停及呼吸不足指数(AHI)和最低氧饱和度差异无显著性意义. 结论自动调定所需压力较低,方法简便,如用于传统CPAP家庭治疗,宜适当上调压力值.