Pages: 1821
Year: Issue:  12
Journal: Journal of Clinical Radiology

Keyword: 
Abstract: 故障现象 意外停电自动关机,来电后开机,控制面板所有灯不亮,扫描床无法移动. 故障分析与检修 根据扫描床无法移动的现象,推测故障为扫描床内供电系统异常所致. 打开扫描床外板,检查安全板,发现安全板电源电压为13V,正常供电电压应为24 V,考虑24 V稳压器出现故障.