Comparative analysis of different ablation catheters treatment on ventriculararrhythmias originating from tricuspid annulus
Author(s): WEN Xutao TAN Haibin YANG Xili XU Zhaoyan HUANG Huanyu
Pages: 383-
386
Year: 2013
Issue:
5
Journal: Journal of Clinical Cardiology
Keyword: 冷盐水灌注导管; 射频消融; 室性心律失常; 非接触心内膜标测; 三尖瓣环; 心电图;
Abstract: 目的:比较不同消融导管治疗三尖瓣环起源室性心律失常的疗效,并分析心电图特征。方法:入选三尖瓣环起源室性心律失常的射频消融患者54例,其中常规消融组21例,冷盐水灌注消融组33例。通过同步12导联心电图分析心电图特征及进行初步定位,然后进行非接触心内膜标测定位消融。结果:与常规消融组比较,冷盐水灌注消融组手术成功率较高(100.0%∶90.5%,P<0.05),手术时间较短[(46.11±7.18)min∶(50.98±10.03)min,P<0.05],放电时间较短[(3.74±1.32)min∶(5.10±2.30)min,P<0.01],放电次数较少[(2.74±1.16)次∶(4.86±2.00)次,P<0.01]。结论:对于三尖瓣环起源室性心律失常的射频消融,冷盐水灌注导管可能有利于提高手术成功率,缩短手术时间,体表发作心电图能对其初步定位提供帮助。