Signal processing in endocardial mapping system
Author(s): Chen Zhong-zhong1, Zhu Jiao-ying1, Su Zhi-jian1, Xia Zhen-hong21School of Mechanical Engineering, Zhengzhou University, Zhengzhou 450001, Henan Province, China, 2Henan Huanan Medical Science and Technology Co., Ltd., Zhengzhou 450001, Henan Province, China
Pages: 4850-
4853
Year: 2011
Issue:
26
Journal: Journal of Clinical Rehabilitative Tissue Engineering Research
Keyword: 心内膜; 标测; 信号处理; 包络检波; 小波降噪;
Abstract: 背景:信号处理是普通导管心内膜三维标测中一项关键技术,需根据其特征设计处理流程,传统处理方法无法消除信号中的大量噪声。目的:进行心内膜三维标测原理的体外模拟实验及信号处理。方法:设计专用压控恒流源做电场激励源以降低噪声,并利用MATLAB信号处理功能对数据进行处理,分别采用带通滤波、包络提取、小波降噪、求均值等方法计算出标测距离,并与实际距离进行对比。结果与结论:该方法有效地降低噪声、直流分量等干扰因素,得到了正确的标测向量,解决了普通导管心内膜三维标测中的信号处理问题,并减小了由呼吸及心跳所引起的定位误差,由此计算所得标测向量误差ε<7%。提示此方法可应用于普通导管心内膜三维标测系统的信号处理。